OCO does not measure CO2 directly; but rather, the intensity of the sunlight reflected from the presence of CO2 in a column of air. This measurement is unique like a fingerprint, and can be used for identification. The instruments use a diffraction grating to separate the incoming sunlight into a spectrum of multiple component colors.

The instruments contain three separate spectrometers that simultaneously measure the intensity of reflected sunlight in three wavelength bands - CO2 (at 1.61 um), CO2 (at 2.06 um) and O2 (at 0.76um). The absorption levels indicate the amount of each gas that the sunlight has passed through.

OCO-3 measurement
Artist interpretation of OCO measurement

The OCO spectra must be very accurate. To eliminate energy from other sources that would generate measurement errors, the light detectors (focal plane assembly) for each spectrometer must remain very cold (-120° C or -184° F). A refrigeration device called a cryocooler is used to maintain and control the focal plane temperature.